Series 1221/1222/1223 Flex-Tube® U-Tube Manometer
Suitable for Pressures up to 100 psi, Better Precision
The Series 1221/1222/1223 Flex-Tube® U-Tube Manometers combine the inherent accuracy of the “U” Tube with the durability of tough, long-lasting plastic construction.The columns are made of 0.375? O.D. flexible and shatterproof clear butyrate tubing and are backed by a white scale channel to provide maximum color contrast. These manometers are ideal wherever a portable, direct reading manometer is needed. See the Support Library tab for Safety Data Sheets.
NIST CALIBRATION available in Product Configurator.
Product Applications
- Duct static pressure
- Calibration labs
- Filter monitoring
Models & Ordering
Features
- Minor Divisions of 1/10" water column (or 2mm water column on metric units) are silkscreened in jet black on stark white scale for positive definition. Easy to read, even at a distance, they are acrylic coated for permanence. All models are furnished with one 3/4 ounce bottle of fluorescein green color concentrate with special wetting agent for W/M models, or 3/4 ounce bottle of red gage fluid for "D" units.
- Series 1221 Manometer - Our simplest, lowest cost basic U-gage. A dependable U-tube manometer that withstands hard use and provides accurate, high visibility readings. For use with water, mercury or red gage fluid. For mercury filled manometers, a scale clamp bar, Dwyer® Part No. A-363 (available as an extra for Series 1221 - and standard on Series 1222) is recommended. One pair of carrying plugs and a pair of non-kink vinyl tube connectors are included with each manometer.
- Series 1222 Manometer - All the features of the 1221 plus magnetic clips for mounting to any vertical steel surface, and clamp bar to insure against U-tube slipping. (Especially recommended for manometers used with mercury.) Both magnets are easily removed and replaced at the user's convenience.
- Series 1223 Manometer - Our finest U-gage - for either portable or stationary use. Safety traps prevent loss of indicating fluid in case of accidental over-pressure. Tubing is permanently bonded to a molded, high impact acrylic plastic top that contains safety traps. Large magnetic clips and clamp bar are provided. Standard type "a" connections include two rapid shut-off type molded nylon tubing connections, two 3' lengths of flexible Tygon® plastic tubing, and two 1/8" pipe thread to tube adapters.
Accessories
Model Number | Description | Usually Ships* | Price | Order | |
---|---|---|---|---|---|
![]() |
A-101 | 1 oz dispenser bottle of red gage fluid, .826 sp. gr. | Same Day | $10.50 | Order |
![]() |
A-102 | 4 oz bottle of red gage fluid, .826 sp. gr. | Same Day | $20.50 | Order |
![]() |
A-103 | 1 pt bottle of red gage fluid, .826 sp. gr. | Same Day | $45.50 | Order |
![]() |
A-104 | 1 qt bottle of red gage fluid, .826 sp. gr. | Same Day | $85.50 | Order |
![]() |
A-126 | 1 oz dispenser bottle fluorescein green color concentrate. | Same Day | $16.00 | Order |
![]() |
A-220 | Clear flexible vinyl tubing is easily inspected, and is therefore best for test applications where a possibility of fluid entering the tubing exists. 3/16" ID x 5/16" OD, lengths to 500'; 45 psi maximum pressure @ 73°F, per ft. | Same Day | $0.50 | Order |
![]() |
A-339 | Adapter, brass, 1/8" NPT to 3/16" rubber and 1/8" ID plastic tubing. | Same Day | $5.40 | Order |
![]() |
A-364 | Magnet assembly for 1222 manometers, 2 required (3 required for 1222-36 and M-1000). | 1-2 Weeks | $10.00 | Order |
![]() |
A-319 | Flexible red PVC connector, 3/16" ID rubber tubing to 1/4" ID plastic tube for 1221,1222 and 1227 manometers. | Less Than 1 Week | $5.70 | Order |
![]() |
A-363 | Scale clamp bar for 1221 manometer. | 1-2 Weeks | $7.50 | Order |
Model Chart
EXAMPLE | 1221 | 8 | W/M | AT | Series 1221-8-W/M-AT U-Tube Manometer, range 4-0-4" w.c. using water/mercury, with aluminum tag. |
SERIES |
1221 1222 1223 |
U-Tube Manometer U-Tube Manometer U-Tube Manometer |
|||
RANGE |
8 12 16 20 24 36 M200 M300 M400 M600 M1000 2000PA 3000PA 4000PA 5000PA 6000PA 9000PA |
4-0-4" w.c. 6-0-6" w.c. 8-0-8" w.c. 10-0-10" w.c. 12-0-12" w.c. 18-0-18" w.c. 100-0-100 mm w.c. 150-0-150 mm w.c. 200-0-200 mm w.c. 300-0-300 mm w.c. 500-0-500 mm w.c. 1000-0-1000 Pa 1500-0-1500 Pa 2000-0-2000 Pa 2500-0-2500 Pa 3000-0-3000 Pa 4500-0-4500 Pa |
|||
FLUID |
W/M D |
Inches of Water/Mercury Inches of Water Using Red Gage Fluid |
|||
OPTIONS |
AT NIST B C D E |
Aluminum Tag NIST Certificate of Calibration Type B Connections (available with 1223 only) Type C Connections (available with 1223 only) Type D Connections (available with 1223 only) Type E Connections (available with 1223 only) |
Product Literature
Manuals
Catalog Pages
SDS
Approvals
Drawings
Questions
- Question
- What are the principle advantages of red gage fluid over water for use in manometers?
- Answer
- Red gage fluid is slow to evaporate, won't freeze, is more readable and produces slightly expanded scales for better reading resolution.
- Question
- Is routine maintenance required for a Dwyer red gage fluid type manometer?
- Answer
- Yes. The zero point should be checked periodically as well as the fluid level. After a year of service, particularly with dirty air, the fluid should be replaced to assure that specific gravity changes are not affecting the accuracy of the reading.
- Question
- Will the water manometers freeze?
- Answer
- Yes, if the temperature drops below 32 degrees Fahrenheit, the fluid in the manometer will freeze. The fluid used in "W/M" style manometers is water-based with a small amount of fluorescent green fluid to enhance the meniscus. The other gage fluids have a lower freezing point, however, it is recommended that the manometers be stored above 32 degrees Fahrenheit.
- Question
- Since water will not mix with oil, why can't a D-style manometer be used to measure water pressures?
- Answer
- The weight of the water itself is also being measured which results in erroneous readings.
- Question
- Why doesn't 1 inch on the scale of a 1223-12-D manometer equal 1 inch on a ruler?
- Answer
- The calibration scale on the manometer has been compensated for the specific gravity of red gage fluid which is less than water.
- Question
- Model 1212. How much fluid mixture is required? What is the best method to fill this unit?
- Answer
- A 1 oz bottle of A-126 Fluorescein green color concentrate is added to a quart of distilled water. Remove a tubing connector from the manometer and pour the mixture in to mid-point level. Shake the manometer to remove air bubbles and slide the scale so the zero mark is in line with the meniscus of the two fluid columns.
Photo Downloads
You must be logged in to view this content.Click here to login to your account.
Product Reviews
Only registered users can write reviews. Please Sign in or create an account